62

Maintenance

9.3.5 Global system view

This option allows you check various components of the instrument:

O2 calibration valve

Gas calibration valve

Reference volume valve

Main purge valve

Needle purge valve

On-off valves function

Regulating valves function

Ultrasound

Flow regulation

Manual displacement

Proportional valve

Check a component by pressing on the button/schematic (which will turn to green) and observe
the values being displayed.

Note:

If the gas flow rate is manually changed, be aware that the automatic purge procedure (every 5

minutes) will reset the gas flow rate back to 0.25 mL/s, which will have an impact on the values being
observed.

9.3.6 System initialization

Selecting this option initiates the standard system startup process, without having to power the
instrument off and on. See

Instrument startup on page 30

for more information on system

startup.

9.3.7 End application

Select this option to perform an orderly shut down of the instrument. On completion, the
instrument can be powered OFF.

9.3.8 Service validation

This option lists the service due date for a number of options. These are user-defined (see
option

Service on page 43

). As each service due date is passed, the Service done button will

be made available. Press this button after the service has been performed to reset the service
due date based on how the service has been configured.

9.3.9 Measurement head initialization

Use this option to return the measurement head to the home position by pressing the Initialize
motor

button.