Yokogawa EJA440E User Manual

Page 85

Advertising
background image

<5. Parameter Summary>

5-7

IM 01C25T01-06EN

Function

Label

Item

Contents

Default value Handling

*

1

Process Alerts Temp Hi Alert Val

High alert value for temperature

Set the threshold value for

high side alarm

120°C (deg C)

W

Temp Lo Alert Val

Low alert value for temperature

Set the threshold value for

low side alarm

-40°C (deg C)

W

Process

variables

Engr Disp

User scaled value

R

Engr Unit

User set engineering unit

Unit specified in Set Engr Unit

8 alphanumerics

R

Loop Current (DD)

AO (DTM)

HART 7

Loop current value

(Analog Output Current)

3.6 to 21.6 mA

R

Loop Current Data

Quality

HART 7

Device variable process data

quality

Device variable process data

quality of Loop Current (mA)

Good

R

Loop Current Limit

Status

HART 7

Device variable limit status

Device variable limit status of

Loop Current (mA)

Not limited

R

PV (DD)

Pres (DTM)

Pressure/Differential pressure

value

R

PV (Pres) Data

Quality

HART 7

Device variable process data

quality

Device variable process data

quality of pressure

Good

R

PV (Pres) Limit

Status

HART 7

Device variable limit status

Device variable limit status of

pressure

Not limited

R

PV Update time

period (DD)

Update time period

(DTM)

HART 7

PV (Pres) Update time period

45 ms

R

% range (DD)

Pres % (DTM)

HART 7

Pressure value in %

-2.50 to 110.00%

R

% rnge (Percent

Range) Data

Quality

HART 7

Device variable process data

quality

Device variable process data

quality of % range (Percent

Range)

Good

R

% rnge (Percent

Range) Limit Status

HART 7

Device variable limit status

Device variable limit status of

% range (Percent Range)

Not limited

R

SV (DD), SP (DTM) Static pressure value

RD

SP %

Static pressure value in %

-10.0 to 100.00 %

RD

SV (SP) Data

Quality

HART 7

Device variable process data

quality

Device variable process data

quality of SV (SP)

Good

R

SV (SP) Limit

Status

HART 7

Device variable limit status

Device variable limit status of

SV (SP)

Not limited

R

SV Update time

period (DD)

SP update time

(DTM)

HART 7

SV (SP) Update time period

360 ms

R

*1:

Handling:

R=Read only, W=Read & Write, M=Method, A=Applicable for option code AL, G=Applicable for option code DG6,

D=Applicable for differential pressure transmitters. Do not change these parameters for pressure transmitters.

*2:

Max three Burst Messages (Burst Message 1, Burst Message 2, Burst Message 3)

*3:

The default value shows MWP (Maximum working pressure) of the capsule.

Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.