Troubleshooting, 1 no flowmeter output under flowing conditions – Yokogawa digitalYEWFLO (DY) User Manual

Page 103

Advertising
background image

IM 1F2B4-01-YIA
Page 98

TROUBLESHOOTING

6.5

F

LOWCHARTS

6.5.1

No flowmeter output under flowing conditions

Yes

No

END

Reset TLA (H02) to '0'

Yes

Yes

No

Are analog/pulse jumpers set

correctly and in agreement with

output type parameter B02?

Apply correct power

No

No

Yes

No

Yes

Output?

END

If remote, is remote

signal cable (YF011)

correctly connected?

END

END

Confirm that B parameters agree

with process conditions, fluid type,

meter size, etc. Correct as required.

Flow may be below Qmin,

check process conditions

& recheck sizing. Correct "B"

parameters as necessary.

Reset TLA (H02) to '0'.

No

No

Yes

Yes

Check TP2 and

COM2, is flowing

waveform present?

A

Is power supplied? Are

voltage and polarity

correct?

Is vortex sensor

correctly connected?

Correct connections

Set TLA(H02) to

-1, Output?

Correct jumbers and/or

parameter