Appendix a: parameter details – Yokogawa digitalYEWFLO (DY) User Manual

Page 118

Advertising
background image

Appendix A: Parameter Details

IM 1F2B4-01-YIA

Page 113

Hand Held

Terminal

Data Range

Indicator/Interface

Help / Remarks

Parameter No. and

Name

Prmtr

No.

Code No.

G22: Min Flow

Read only

No access

Minimum flow reading since the last execute of G20.
Display value 0 - 110% span.

G60: Self-check

Good
Error
Read only

No access

Diagnostic self-check. If good is displayed no error
exists. See “How To Use Diagnostics” for details.

H: Maintenance

Read only

No access

This signifies the beginning of the Maintenance Menu.

H01: N. Balance

-5 to 10
Factory Set

H1

-5 to 10

Noise Balance:
This parameter is factory set to match the sensor
crystals for maximum signal to noise ratio. Field
adjustment should be done only by a trained
technician, or as directed by the Factory. Refer to
“Troubleshooting” for more details.

H02: TLA

-1
0
1
2
default is 0

H2

Trigger Level Adjustment: Factory default is 0
This adjustment is used to suppress the effects of pipe
vibration noise on the output signal. -1 is very sensitive
and 2. Refer to “Troubleshooting” for more details

H03: Gain

16 steps
default is 0

H3

16 steps

Amplifier Gain: Factory default is 0
Gain is set automatically by values input in the B
menu. Consult the factory prior to adjusting this
parameter, or refer to “Troubleshooting” for more
details.

H04: H. F. Filter

16 steps
default is 0

H4

16 steps

High Frequency Filter: Factory default is 0
Filter is set automatically by values input in the B
menu. This parameter is used to high frequency noise.
Consult the factory prior to adjusting this parameter, or
refer to “Troubleshooting” for more details.

H06: Noise Judge

Not active
Active

H6

0
1

Dynamic Noise Reduction:
This parameter determines if the input received is
noise or flow signal by comparing amplitude and
frequency to predicted values. Parameter B08 Min.
Density Flowing adjusts the cutoff value of this
function up and down. When injecting a frequency
signal into TP2 for amplifier calibration this parameter
must be in the Not Active mode. Refer to
“Calibration” and “Troubleshooting” for details.

H07: L.C. Flowrate

0.00001 to B52 value

H7

0.00001 to B52

This parameter sets the low flow cut off value. Any
signal below this value will be ignored. Refer to “How
to set the Low Flow cutoff value” for details.

H08: Trim 4 mA

-1.0000 to 10.000 %

No access

This parameter is used for fine adjustment of the 4 mA
output. Refer to “How to trim the 4-20 mA output” for
more details.

H09: Trim 20 MA

-10.000 to 10.000 %

No access

This parameter is used for fine adjustment of the 20
mA output. Refer to “How to trim the 4-20 mA output”
for more details.

H10: CLR. Err2

Not active
Active

No access

ERR #2 on the local indicator, warns that the span value
in B52 exceeds 32.8 ft/sec for liquids, or 262 ft/sec for
gas or steam. Select Active to disable this warning.

H20: Measure TP2

Not execute
Execute

No access

Select Execute to allow digital display of the TP2 test
point voltage in parameter H21 (Volts, peak-to-peak).