Troubleshooting – Yokogawa digitalYEWFLO (DY) User Manual

Page 107

Advertising
background image

IM 1F2B4-01-YIA
Page 102

TROUBLESHOOTING

6.5.4

Output is unstable when flowrate is low

YES

NO

Perform piping checkout

procedure, section 6.4.3.

Perform noise balance adjustment,

section 6.4.4.

Operation

Normal?

END

NO

YES

YES

NO

Is TP2-COM2

waveform unstable?

Confirm that vortex sensor

assembly mounting bolts

are properly torqued.

Check vortex

shedder for obstructions

Check meter sizing. Is
the flowrate within the

measurable range?

Resize meter

Waveform

superimposed with

60Hz noise?

YES

NO

END

Eliminate noise by grounding, or

using shielded wire. For remote meters

only, confirm that Yokogawa's

special signal cable (YF011) is used

and properly terminated.

Contact Yokogawa Industrial Automation

technical assistance, 800/524-7378.