Appendix A: Parameter Details

IM 1F2B4-01-YIA
Page 112

Hand Held

Terminal

Data Range

Indicator/Interface

Help / Remarks

Parameter No. and

Name

Prmtr

No.

Code No.

D60: Self-check

Good
Error
Read only

No access

Diagnostic self-check. If good is displayed no error
exists. See “How To Use Diagnostics” for details.

E: Control

Read only

No access

This signifies the beginning of Menu E for
configuration of the local indicator.

E01: Total Reset

Not execute
Execute

E1

0
1

Select Execute to reset the integral totalizer.

E02: Disp Select

Rate (%)
Rate
Total
Rate (%), Total
Rate , Total
Rate, Rate (%)

E2

0
1
2
3
4
5

Select the desired display mode for the list.

Comma separated items will alternate every 4 sec.

E60: Self-check

Good
Error
Read only

No access

Diagnostic self-check. If good is displayed no error
exists. See “How To Use Diagnostics” for details.

F: Test

Read only

No access

This signifies the beginning of Menu F. For loop
testing purposes.

F01: Out Analog

0.00001 to 110.00 %

F1

0.00001 to 110.00 To check the operation of the analog output loop, input

a value between 0 and 110%. The current output and
the Rate% display will respond to this input.
Engineering unit display and totalization will continue
to measure the true flow reading.

F02: Out Pulse

0.00001 to 6000 Hz

F2

0.00001 to 6000

To check the operation of the pulse output circuit,
input a frequency value between 0 and 6000Hz. The
pulse output will respond to this input. All display and
totalization functions will continue to measure the true
flow reading.

F60: Self-check

Good
Error
Read only

No access

Diagnostic self-check. If good is displayed no error
exists. See “How To Use Diagnostics” for details.

G: Check Data

Read only

No access

This signifies the beginning of Menu G. Real time
information regarding the process will be displayed
within this menu.

G01: Frequency

Read only

No access

Displays actual frequency of vortex shedding, in Hz

G02: Span Freq

Read only

No access

Displays calculated span frequency, in Hz, which
corresponds to the 20 mA value. Including any effects
of correction factors in the D menu.

G03: Velocity

Read only

No access

Displays velocity of the fluid in m/s.
1 m/s = 3.28 ft/sec

G04: Span Vel

Read only

No access

Displays velocity, in m/s, which corresponds to the
span, or 20 mA value, of the meter.
1 m/s = 3.28 ft/sec

G05: Reynolds

Read only

No access

Displays 1/1000 of the Reynolds number
corresponding to the measured flow rate.

G20: Fluctuation

Not execute
Execute

No access

Executing this parameter resets and starts the
collection of Max and Min flow data for display in
parameter G21 (Max) and G22 (Min).

G21: Max Flow

Read only

No access

Maximum flow reading since the last execute of G20.
Display value 0 - 110% span.