Troubleshooting, Vi. error codes 6.1 e – Yokogawa digitalYEWFLO (DY) User Manual

Page 93

Advertising
background image

IM 1F2B4-01-YIA
Page 88

TROUBLESHOOTING

VI.

ERROR CODES

6.1

E

RROR

C

ODE

L

ISTING

Effect on Output

Effect on Indication via local Digital Display

Error

No.

Diagnostic

Message

Error definition and

Probability Cause

Current

Output

Pulse Output

Rate in %

Output

Rate in

Engineering

Units

Totalizer

Corrective Action

1

Over

Output

This error signifies that the

flow exceeds the meter span

by over 110%.

Output driven

to a fixed

110%.

Normal

Operation

Indication is

driven to a

fixed 110% of

the span.

Normal

Operation

Normal

Operation

Respan the meter or

correct the overrange

condition.

2

Span Set

error

A span setting error

indicates that the span

velocity limits have been

exceeded. (32 ft/sec for

liquids and 262 ft/sec for

gas). Sensor may be

damaged. Contact

Yokogawa Industrial

Automation service

department for analysis.

Output may be

non-linear.

Output may be

non-linear.

Output may be

non-linear.

Output may be

non-linear.

Output may be

non-linear.

Correct the span

setting in parameter

B51 and B52 to meet

the velocity

limitations of the

meter.

3

NJ circuit

error

Noise discrimination circuit

is not functioning properly.

Normal

Operation

Normal

Operation

Normal

Operation

Normal

Operation

Normal

Operation

Amplifier replacement

required. Contact

Yokogawa Industrial

Automation service

dept.

4

Pulse Out

error

The pulse output span has

exceeded 6K Hertz.

Normal

Operation

Output is

limited to 6K

Hertz.

Normal

Operation

Normal

Operation

Normal

Operation

Check parameter

settings, in particular

C02.

5

EE Prom

error

EE prom is not functioning

correctly.

Output is fixed

at 1.25%.

No output

Output is fixed

at -1.25%.

Output is fixed

at 0.0%.

No output

Amplifier replacement

required. Contact

Yokogawa Industrial

Automation service

dept.

---

CPU error

The CPU has failed and the

meter is not functioning.

The display and self-

diagnostics are completely

inoperable.

Output is fixed

at 1.25%.

No output

No output

No output

No output

Amplifier replacement

required. Contact

Yokogawa Industrial

Automation service

dept.

Advertising