Appendix A: Parameter Details

IM 1F2B4-01-YIA

Page 111

Hand Held

Terminal

Data Range

Indicator/Interface

Help / Remarks

Parameter No. and

Name

Prmtr

No.

Code No.

E - 1
UNSC*1
UNSC*10
UNSC*100

10
11
12
13

E - 1 = 0.1 (10

-1

) flow units per pulse

UNSC *1 = K factor x 1 = 1 pulse out/pulse in
UNSC*10 = K factor x 10 = 10 pulses out/pulse in
UNSC*100 = K factor x 100 = 100 pulses out/pulse in

C09: Unit Conv. Fa

0.00001 to 32000

C9

0.00001 to 32000 Use when custom flow units are needed.

Value = (std. span units/custom unit)
Refer to "How to set up for user defined flow units" for
details.

C10: Users Unit

8 characters

No access

Using the Alphanumeric keys enter the name of the
custom flow unit. Only the first 3 characters will be
displayed in Parameter: A20.

C60: Self-check

Good
Error
Read only

No access

Diagnostic self-check. If good is displayed no error
exists. See “How To Use Diagnostics” for details.

D: Adjust

Read only

No access

This signifies the beginning of the D Menu. This Menu
allows access to special compensation features.

D01: Reynolds Adj

Not active
Active

d1

0
1

This parameter is used to improved accuracy when
measuring flow between a Reynolds No. Of 5,000 and
20,000. Accurate viscosity and density values must be
provided. Refer to “How to activate Reynolds Number
correction” for details.

D02: Viscosity

0.00001 to 32000

d 2

0.00001 to 32000 Enter operating viscosity in centipoise, for use with

Reynolds number correction only.

D03: Density f

0.00001 to 32000

d 3

0.00001 to 32000 Enter operating density units per B07, for use with

Reynolds Number correction only.

D05: Pipe Effect

Not active
Wafer Sch 10
Wafer Sch 40
Wafer Sch 80
Flange Sch 10
Flange Sch 40
Flange Sch 80

d 5

0
1
2
3
4
5
6

This function corrects for errors that may be caused
when YEWFLO is installed in pipe other than Sch 40
Pipe. To invoke this function, select the connection
and pipe Sch from the list. Refer to “How to activate
Mis-matched pipe schedule correction” for details.

D06: Expansion FA

Not active
Active

d 6

0
1

This function improves gas accuracy above 35 m/s to
0.8% of rate, when metering a compressible fluid (Gas
or Steam). To activate this function, select ACTIVE.
Refer to “ How to increase gas and steam accuracy by
correcting for gas expansion” for details.

D20: Flow Adjust

Not active
Active

No access

Select active to activate this linearization feature and
input the appropriate data in parameters D21 through
D30. The correction data for this function requires
special calibration. Call Factory for details.

D21: Freq 1

0.00001 to 32000

No access

First Break Point Frequency (Hz)

D22: Data 1

-50.00 to 50.00

No access

First Break Point Correction (%)

D23: Freq 2

0.00001 to 32000

No access

Second Break Point Frequency (Hz)

D24: Data 2

-50.00 to 50.00

No access

Second Break Point Correction (%)

D25: Freq 3

0.00001 to 32000

No access

Third Break Point Frequency (Hz)

D26: Data 3

-50.00 to 50.00

No access

Third Break Point Correction (%)

D27: Freq 4

0.00001 to 32000

No access

Fourth Break Point Frequency (Hz)

D28: Data 4

-50.00 to 50.00

No access

Fourth Break Point Correction (%)

D29: Freq 5

0.00001 to 32000

No access

Fifth Break Point Frequency (Hz)

D30: Data 5

-50.00 to 50.00

No access

Fifth Break Point Correction (%)