Troubleshooting – Yokogawa digitalYEWFLO (DY) User Manual

Page 104

Advertising
background image

TROUBLESHOOTING

IM 1F2B4-01-YIA

Page 99

No flowmeter output under flowing conditions (cont.)

Yes

END

Output?

END

Upload parameters to

handheld terminal

Install spare amp

Output?

END

OR

Download saved parameters

to new amp

Go to A,

"Check TP2 and COM2"

Reduce minimum density

(B08) value by 20%.

Yes

Yes

No

No

No

Check and reset

Gain = 0 (H03),

Output?

Set noise judge

(H06) to not active,

Output?

Check and correct minimum

density parameter (B08) to

process conditions. Set

noise judge (H06) to active.

Reset minimum density

to original setting

No

Yes

Yes

No

Less than or equal to

10 loops?

Contact Yokogawa Industrial Automation

technical assistance, 800/524-7378