Maintenance – Yokogawa digitalYEWFLO (DY) User Manual

Page 66

Advertising
background image

MAINTENANCE

IM 1F2B4-01-YIA
Page 70

4.1.12 Setting up the local LCD indicator display mode

The YEWFLO vortex meter offers a variety of display options. The following procedure will allow you to
select which values are displayed, for your application.

BT200 handheld terminal

BT100 handheld terminal

E: CONTROL

Scroll through the menus to
the E: CONTROL menu, then
press ENTER.

Press the MENU key until
the E: Control menu is
displayed.

E02: DISP SELECT
rate %

Scroll to parameter E02: DISP
SELECT then press ENTER.

Using the p r m tr key
move to parameter E02: Disp
Select .

Scroll to your desired choice,
press ENTER twice.

Using INC or DEC move
to the display option of your
choice, press ENT twice to
select.

Select one of the 6 display options for the local indicator.
The available options are:
rate %

Rate in percent of span.

rate

Rate in engineering units.

total

Totalized flow in engineering
units.

rate%, total

Alternating display; rate in
percent of span and total flow.

rate, total

Alternating display; rate in
engineering units total flow.

rate, rate%

Alternating display; rate in
engineering units and rate in
percent of span.

Advertising