MAINTENANCE

IM 1F2B4-01-YIA
Page 72

4.1.14 Trimming the 4-20 mA analog output

The 4-20 mA analog output circuit is accurately calibrated at the factory using precision test equipment.
Field adjustment of this circuit is rarely required. This adjustment should not be made when there is a
suspected offset of the 4 mA point due to noise such as pipe line vibration, in this case, please refer to
signal conditioning.

If you find it necessary to trim the 4 mA and/or 20 mA value, it can be adjusted digitally with the
handheld terminal. An example of when to use this adjustment would be when there is a 250

resistor in

the loop changing the output to 1 - 5 volts. If the resistor value is not exact, an offset error of either the 4
mA (1V) and/or 20 mA (5V) reading will occur.

To digitally trim the current output, refer to the following formula and example to determine the correct
digital trim parameter settings. The parameter H08 :Trim 4mA is used to adjust the 4 mA (or 0%) output,
and parameter H09 :Trim 20mA is used to adjust the 20 mA (or 100%)

H :Maintenance

BT200 handheld terminal

BT100 handheld terminal

H08 :TRIM 4ma
0.45 %

Move to parameter H08, re-enter the
value shown, (.45 this example) press
Enter twice. The output will go to
0%. (Don’t press OK until you are ready
to leave this parameter).

Move to parameter H08 and press e

nt

twice.

Read the analog output at 0% and record it as X (3.93mA this example).

Record the original value of H08 (0.45 this example).

H09 :Trim 20ma
1.25 %

Move to parameter H09 and re-enter the
value shown, (1.25 this example) press
enter twice. The output will go to
100%. (Don’t press OK until you are
ready to leave this parameter).

Move to parameter H09 :TRIM 20ma and
press e

nt twice.

Read the analog output at 100% and record it as Y (19.72mA this example).

Record the original value of H09 (1.25 this example).

Calculate the New H08 setting using X and Y from above in Equation 1.

Calculate the New H09 setting using X and Y from above in Equation 2.

H08 :TRIM 4ma
0.8993 %

Store the New H08 setting (confirm the output goes to 4 mA or 0%).

H09 :Trim 20ma
2.58%

Store the New H09 setting (confirm the output goes to 20 mA or 100%).