3. model and suffix codes, 3. model and suffix codes -3 – Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 10

Advertising
background image

IM 12D08E02-01E

Specification 2-3

E. Housing

: Cast aluminium case

with chemically resistant

coating, cover with flexible

polycarbonate window. Case

color is off-white and cover is

moss green. Cable entry is via

six PG13.5 nylon glands. Cable

terminals are provided for up to

2.5 mm finished wires. Weather

resistant to IP65. Pipe wall or

panel mounting, using optional

hardware.

F. Data protection

: EEPROM for configuration and

logbook, and lithium battery for

clock.

G. Watchdog timer

: Checks microprocessor

H. Automatic safeguard

: Return to measuring mode

when no keystroke is made for

10 min.

I. Power interruption

: Less than 50 milliseconds no

effect.

J. Operation protection

: 3-digit programmable pass-

word.

K. Safety and EMC conforming standards

,

Safety

: conforms to EN 61010-1,

EMC

: EN 61326-1 Class A, Table 2

(For use in industrial locations) (Note 1)

EN 61326-2-3

EN 61000-3-2 Class A

EN 61000-3-3

EMC Regulatory Arrangement in

Australia and New Zealand (RCM)

EN 55011 Class A, Group 1

Korea Electromagnetic

Conformity Standard Class A

한국 전자파적합성 기준

Installation altitude: 2000 m or less

Category based on IEC 61010: II (Note 2)

Pollution degree based on IEC 61010: 2

(Note 2)

Note 1: This instrument is a Class A

product, and it is designed for

use in the industrial environment.

Please use this instrument in the

industrial environment only.

2-3. Model and suffix codes

[Style: S2]

Model

Suffix
code

Option
code

Description

DC402G

-------------

----------

Dual Conductivity

Converter

Type

-1

----------

General

Power Supply

Voltage

-1

-2

----------

----------

115V +/-15% AC, 50/60 Hz

230V +/-15% AC, 50/60 Hz

Language

-E

-J

----------

----------

English

Japanese

Options

Mounting Hardware

Hood

Tag Plate

Conduit Adapter

/U

/PM

/H3

/H4

/SCT

/AFTG

/ANSI

/X1

Pipe, wall mounting

bracket (Stainless steel)

Panel mounting bracket

(Stainless steel)

Hood for sun protection

(Carbon steel)

Hood for sun protection

(Stainless steel)

Stainless steel tag plate

G 1/2

1/2 NPT

Epoxy baked finish (*1)

*1 The housing is coated with epoxy resin.

A급 기기 (업무용 방송통신기자재)
 이 기기는 업무용(A급) 전자파적합기기로서 판매자 또는
사용자는 이 점을 주의하시기 바라며, 가정외의 지역에서
사용하는 것을 목적으로 합니다.

Note 2: Installation category, called over-

voltage category, specifies impulse

withstand voltage. Category II is

for electrical equipment.

Pollution degree indicates the

degree of existence of solid, liquid,

gas or other inclusions which

may reduce dielectric strength.

Degree 2 is the normal indoor

environment.

Advertising