3. wiring the power supply, 3-1. general precautions, 3-2. access to terminal and cable entry – Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 16: 3. wiring the power supply -5, 3-1. general precautions -5, 3-2. access to terminal and cable entry -5

Advertising
background image

IM 12D08E02-01E

Installation and wiring 3-5

FRONT GLANDS

REAR GLANDS

Sensors

0/4-20 mA Output

signals

Contact

output

Contact

output

Contact

input

0/4-20 mA

Power

S1

S2

S3

S4/FAIL

Figure 3-6. System configuration

3-3. Wiring the power supply

3-3-1. General precautions

Make sure the power supply is switched off. Also, make sure that the power supply is correct for the

specifications of the EXA and that the supply agrees with the voltage specified on the nameplate.

Remove the front cover by unscrewing the four screws to check this nameplate on the top of the display

board.

Local health and safety regulations may require an external circuit breaker to be installed. The instru-

ment is protected internally by a fuse. The fuse rating is dependent on the supply to the instrument. The

250 VAC fuses should be of the “time-lag” type, conforming to IEC60127.

The internal fuse is located next to the power terminals (in the lower right hand corner).

Use only a fuse of the specified current, voltage and type ratings to prevent fire. For fuse replacement,

refer to Section 7-3, “Fuse Replacement.”

3-3-2. Access to terminal and cable entry

Terminals 1 and 2 on the bottom terminal strip are used for the power supply. Guide the power cables

through the gland closest to the power supply terminals. The terminals will accept wires of 2.5 mm

2

(14

AWG). Use cable finishings if possible.

Connect the wires as indicated in the wiring diagram (refer to figure 3-6).