Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 76

Advertising
background image

IM 12D08E02-01E

Appendix 2 11-3

3/3

QIS 12D08E02-01E

3.5 Temperature

Indication Check

Following Section 3.4, press the [ENT] key until the message display shows “PT1000.” In

this state, change the resistance of the decade resistance box 1 and check the data

display. The value on the data display must be within the range shown in Table 3.

Table 3

Temperature (°C)

Resistance () of

Decade Resistance Box 1

Data Display (°C)

–10

960.9

–10 ±0.3

25

1097.3

25 ±0.3

190

1721.6

190 ±0.3

3.6 Resistance Indication Check

Following Section 3.5, press the [ENT] key until the message display shows “RES 1.”

Each time the [ENT] key is pressed, the message display changes : “RES 1”  “RES 2”

“RES 3” ……  “RES 12.” When the message display shows the items in Table 4, set

the resistance of the decade resistance box 2 to the corresponding value in Table 4.

Check the data display and the value must be within the range shown in Table 4.

Continue to press the [ENT] key to return to normal measurement mode.

Table 4

Message Display

Resistance (Ω) of

Decade Resistance Box 2

Data Display (Ω)

RES3

160

160 ±0.8

RES5

800

800 ±4

RES7

3.2k

3.2 ±0.02k

RES9

16k

16 ±0.08k

RES11

64k

64 ±0.4k

Figure 1 Testing Circuit and Test Equipment

DC402G

11 12 14 15 11 12 14 15 66 65

62

61

22

21

71

72

51

52

41

42 31 32 3 2

1

+

-

+

-

SENSOR 1

SENSOR 2

mA2 mA1

Decade

Resistance

Box 1

Decade

Resistance

Box 2

Power

Supply

-

+

300

DC

Ammeter

-

+

300

DC

Ammeter