10-9. WHAT IS DUAL CONDUCTIVITY?

Dual cell conductivity is a precise, comparative

measurement.

The EXA DC402G receives inputs from two

conductivity cells located at different points in

the process and compares them according to

one of six programmed user-selectable formula

- Ratio (a/b)

- Differential or linear difference (a-b)

- Percent passage (b/ax 100)

- Percent rejection {(a-b)/a x100}

- Deviation {(b/a) a x100}

- pH calculation acc. VGB directive 450L

The output signal corresponds directly to the

formulas. The EXA DC402G also displays the

absolute value of each cell on a second display

line, as desired by the user.

The unit displays all values in conductivity units

(µS/cm or mS/cm), percentage (%), resistivity

(MΩ•cm) or pH.

The pH value of the (boiler) water/condensate

is calculated from the conductivity value before

and after the cation filter according the VGB

directive 450L.

12

11

10

9

8

7

0.1

1.0

10

100

pH

SC(before) - SC(cation) / 3

pH as function of differential Conductivity

This pH calculation requires two conductivity

values which are compensated to 25ºC.

Normally matrix compensation for morpholine/

ammonia and cation are used (service code 22

should be set to 2.1 or 3.1). Consequently the

pH value displayed corresponds to 25ºC.

RETURN LINE

COOLING

TOWER

DUAL CELL

CONDUCTIVITY

RATIO

A/B

B

A

MAKE-UP WATER

COOLING WATER SYSTEM

BLOW DOWN LINE

Fig. 10-2. Ratio output controls blow-down of cooling tower

based on concentration factor.

A

B

DIFFERENTIAL A-B

WASHING PROCESS

Fig. 10-3. Differential output water flow to optimize washing

efficiency.

Fig. 10-5. pH output as a function of two conductivity

values before and after the cation exchanger.

Fig. 10-4. Deviation output alarms directly after leakage in the

heat exchanger.

A

B-A X 100

(TA - TB)

SEA WATER

CONDENSATE

A

B

SEA WATER

DEVIATION

DIFFERENTIAL

TEMPERATURE

HEAT EXCHANGER

Advertising