Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 40

Advertising
background image

IM 12D08E02-01E

Parameter setting 5-15

Code Display

Function

Function detail

X

Y

Z

Default values

Parameter specific functions
01

*SC.RES

Select main parameter Conductivity

0

0

Cond.

Resistivity

1

02

Not used

03

*CC1/

Set cell constant

Press NO to step through choice of

0.100 cm

-1

*CC2

multiplying factors on the second display.

0.10xC

0.10xC

1.00xC

10.0xC

100.xC

0.01xC

Press YES to select a factor

Use >, ^, ENT keys to adjust MAIN digits

1.000

04

*AIR 1/*AIR 2 Zero calibration

Zero calibration with dry cell connected

*START

Press YES to confirm selection

*”WAIT”

Press YES to start, after briefly displaying

*END

“WAIT”, *END will be displayed

Press YES to complete

05

*POL.CK

Polarization check

Polarization check off

0

Polarization check on

1

1

On

06-09

Not used

Code Display

Function

Function detail

X

Y

Z

Default values

Temperature measuring functions
10

*T.SENS

Temperature sensor Pt1000

0

0

Pt1000

Ni100

1

PB36

2

Pt100

3

8k55

4

11

*T.UNIT

Display in °C or °F

°C

0

0

°C

°F

1

12

*T.ADJ 1

Calibrate temperature Adjust reading to allow for cable

None

*T.ADJ 2

resistance.

Use >, ^ , ENT keys to adjust value

13-19

Not used

Code 12 *T.ADJ 1

First select sensor 1 or sensor 2 for temperature adjustment

(*T.ADJ 1 or *T. ADJ

2)

.

With the process temperature sensor at a stable known temperature, the

temperature reading is adjusted in the main display to correspond.

The calibration is a zero adjustment to allow for the cable resistance, which will

obviously vary with length.

*T.ADJ 2

The normal method is to immerse the sensor in a vessel with water in it,

measure the temperature with an accurate thermometer, and adjust the reading

for agreement.