Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 75

Advertising
background image

IM 12D08E02-01E

11-2 Appendix 2

2/3

QIS 12D08E02-01E

e. Press the [>] key once. The data display will show “80” with the second digit of 0

flashing.

f. Press the [

] key seven times. The data display will show “87” with the second

digit of 7 flashing.

g. Press the [ENT] key. The message display will show “PASS.”

(2) Setting Password 070.

a. Press the [>] key once. The data display will show “000” with the second digit of 0

flashing.

b. Press the [

] key seven times. The data display will show “070” with the second

digit of 7 flashing.

c. Press the [ENT] key. The message display will show “HIF.”

d. Press the [YES] key.

(3) Confirming the date and time.

a. Press the [ENT] key. The data display will show the date in day, month, year

format.

b. Press the [ENT] key. The data display will show the time in hour, minute, second

format.

c. Press the [ENT] key. The message display will show “mA 1+2.”

This is the inspection mode. When the message display shows “mA 1+2,” check the

current outputs.

Each time the [ENT] key is pressed, the value on the data display increase in steps of 4.0

like “4.0,” “8.0,” and “12.0.” Check the current outputs 1 and 2. Corresponding to the data

display, the current output must be within the range shown in Table 1.
When the data display shows “12.0,” pressing the [ENT] key causes the message display

to show “RIPPLE” and the data display to remain unchanged. Press the [ENT] key again.

The message display will show “mA 1+2” and the data display “16.0.”

Table 1

Data Display

Current Output (mA DC)

4.0

4 ±0.02

12.0

12 ±0.02

20.0

20 ±0.02

3.4 Contact Operation Test

Following Section 3.3, press the [ENT] key until the message display shows “JUMPER.”

Then press the [ENT] key once, the message display will show “WAIT” and LEDs S1,

S2, WASH/S3, and FAIL/S4 flash rapidly. Check that the message display shows and

LEDs turns on/off in the order as shown in Table 2 and that a message of “FAIL x” does

not appear.

Table 2

Message

Display

S1 S2 S3

FAIL/S4

LED1 OFF ON ON OFF
LED2 ON OFF ON OFF
LED3 ON ON OFF OFF
LED4 ON ON ON ON

Advertising