IM 12D08E02-01E

Installation and wiring 3-9

3-8. Other sensor systems

To connect other sensor systems, follow the general pattern of the terminal connections as listed below:

11 and 12

Always used for temperature compensation resistor input (Pt1000, Ni100, Pt100, PB36

and 8k55)

14

Normally used for the outer electrode

15

Used for inner electrode

In case a 4-electrode measuring system will be used, 14 and 16 should be used for the current elec-

trodes.Please ensure that shielded cabling will be used.

In figure 3-10 this is shown in a schematic way.

Figure 3-10. Connection diagram for other sensors

Figure 3-11. Terminal identification labels example

t

11 12

TEMPERATURE

SENSOR

CELL

ELECTRODE

14 15

2-electrode configuration

23

SCREEN

mA1

DC402 REFER TO INSTRUCTION MANUAL FOR CONNECTIONS

mA2

SCREEN

mA OUTPUT

CONT

SENSOR 1

22 21 11 12 14 15 11 12

63 66 65 62 61 95 94 93 92 91

14 15

SENSOR 2

S1

NO

NC

C

S2

NO

NC

C

S3

NO

NC

C

1

2

3

33

32

31

43

42

41

53

52

51

73

72

71

S4

NO

NC

C

L

N

G

FUSE

250V

T200mA

115 VAC

250VAC

5A

100VA

250VDC

5A

50W

Sensor Inputs

Relay Contacts

Power Supply

mA Outputs