7. configuration checklist for dc402g, 7. configuration checklist for dc402g -6 – Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 69

Advertising
background image

IM 12D08E02-01E

10-6 Appendix

10-7. Configuration checklist for DC402G

Standard Configuration

Options

Reference for

change

Measured Variable(s)
primary inputs

Conductivity (SC) and Temp Resistivity i.s.o Conductivity

code 01

conductivity range

0.000 µS/cm - 100 µS/cm

any span within 0.000µS/cm - 1999mS/cm

“range”

conductivity units

Auto ranging µS/cm - mS/cm Choice out of fixed µS/cm or mS/cm

code 56

resistivity range

0 - 19.99 MΩ•cm

any span within 0.000kΩ•cm - 999MΩ•cm

“range”

resistivity units

MΩ•cm

Auto ranging or other fixed values

code 56

temperature range

0 - 100 °C

any span in -20 ... +250

“range”

temperature unit

Celsius

Fahrenheit

code 11

Outputs

analog output

4- 20 mA for SC

0- 20 mA or 4- 20 mA

code 30

second output

4- 20 mA for Temp

0- 20 mA or 4- 20 mA

code 30

output allocation

SC and Temp

SC, Resistivity, Temp, Table, PI control

code 31

contact outputs

S1= high at 100 µS/cm

(4) freely programmable

“setpoint”

S2= low at 100 µS/cm

code 40. 41, 42, 43

S4= FAIL

contact allocation

mS/cm and FAIL

µS/cm, mS/cm, kΩ•cm, MΩ•cm,

code 40- 43

temp, PI control, HOLD, FAIL

contact variables

dead time= 0.2 s; hyst= 0.1% time: 0- 200 s; hyst 0.1- 100%

code 44

add. contact functions

none

time out alarm

code 47

control functions

none

PI on contacts or mA output

code 45, 46, 34, 33

digital outputs

none

RS485

code 60

Communication

digital interface

disabled

RS485

code 60

variables on display

µS/cm (mS/cm) and temp

TC, %w/w, °C, mA1, mA2, CC, TR, REL

“display”

burn out

disabled

burn low (3.5)/ high (22) on mA1/ mA2

code 32

password protection

disabled

for maint/ comm./ serv level

code 52

autoreturn

return to measure in 10 min. enable or disable

code 50

add. function in MAINT

disabled

setpoint adj

code 51

Diagnostics

check on polarization

enabled

enable or disable

code 05

check on cell constant

active

“calibrate”

check on TC coefficient

active

“temp.”

Compatibility

SC sensor

SC40, SX42

SC8SG, L&N compatible

code 10

temperature sensor

Pt1000

Ni100, PB36, Pt100, 8k55

code 10

cell constant

0.100/cm

from 0.008/cm up to 50.0/cm

code 03

Special Features

temperature calibration

none

adjustment +/- 15 °C

code 12

zero calibration

none

adjustment -1 µS/cm

code 04

temp. comp.

acc to NaCl tables IEC 60746-3 NaCl, manual TC, matrix

“temp.”,

code 20 - 28

HOLD during maintenance

disabled

hold last or hold fix

“Hold”

contact during HOLD

disabled

possible on S1,S2 or S3

code 40- 42

soft fail alarm

disabled

possible for E1, E5...E13, E22

code 53