3-3-3. AC power

Connect terminal 1 to the phase line of the AC power and terminal 2 to the zero line. The size of con-

ductors should be at least 1.25 mm

2

. The overall cable diameter should be between 6 & 12 mm (0.24 &

0.47 in).

3-3-4. Grounding the housing

Protective grounding must be made to prevent electric shock.

To protect the instrument against interference, the housing should be connected to ground by a large

area conductor. This cable can be fixed to the rear of the housing using a braided wire cable.

See figure 3-8.

Please be sure to connect protective grounding of

DC402G with cable of 1.25 mm

2

or larger cross sec-

tion in order to avoid the electrical shock to the opera-

tors and maintenance engineers and prevent the

influence of external noise. And further connect the

grounding wire to the mark (100Ω or less).
3-3-5. Switching on the instrument

After all connections are made and checked, the

power can be switched on from the power supply.

Make sure the LCD display comes on. All segments

will illuminate, then the instrument will momentarily

display its unique serial number. After a brief inter-

val, the display will change to the measured value. If

errors are displayed or a valid measured value is not

shown, consult the troubleshooting section (Chapter

8) before calling Yokogawa.

Figure

3-8.

Grounding

the

housing

DANGER

CAUTION

71

S4

S3

S2

S1

C NC NO

72 73 51 52 53 41

43 31

33

42

32

250VAC

5A

100VA

250VDC

5A

50W

FUSE

115

230

250VAC; T 200mA

3

1

2

G N

L

C NC NO C NC NO C NC NO

VAC

VAC

12

11

14 15

22 21

63 66 65 62 61 95 94 93 92 91

Sensor inputs

SENSOR 1

12

11

14 15

mA Outputs

CONT

SENSOR 2

mA OUTPUT

Relay Contacts

Power Supply

Temp

Temp

mA2

mA1

Screen

250VAC; T 100mA

DC402 REFER TO INSTRUCTION MANUAL FOR CONNECTIONS

23

Screen

(M4 screw)

Advertising