3. fuse replacement, 3. fuse replacement -2 – Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 59

Advertising
background image

IM 12D08E02-01E

Maintenance 7-2

7-3. Fuse Replacement

This section describes how to replace the power supply fuse.

• Use only a fuse of the specified current, voltage and type ratings to prevent fire.

• Power to the instrument must be turned off at the external breaker before replacing the fuse. After

replacing the fuse, install a high-voltage shield plate before turning the power on.

• Do not short circuit the fuse holder.

Specified Ratings of the Power Supply Fuse

The power supply fuse (recommended fuse) used in the DC402G is as follows.

• 115 V AC version

Maximum rated voltage: 250 V

Maximum rated current: 200 mA

Compliance:

UL, CSA, VDE, Japan’s Electrical Appliance and material Safety Law

Part number:

A1105EF

• 230 V AC version

Maximum rated voltage: 250 V

Maximum rated current: 100 mA

Compliance:

UL, CSA, VDE, Japan’s Electrical Appliance and material Safety Law

Part number:

A1103EF

How to replace the fuse

1) Before replacing the fuse, turn off power to the instrument at the external breaker.

2) Remove the instrument cover and the high-voltage shield plate.

3) Remove the cover from the fuse holder by pulling out by hand.

4) Remove the fuse and install a new, recommended or equivalent fuse on the holder.

5) Put the fuse cover back on the holder securely.

6) Install the high-voltage shield plate.

It is recommended that the fuse be replaced every two years even if it has not blown.

Shield Plate

Fuse Cover

Fuse Holder

Figure 7-1. How to replace the fuse