2. operating specifications, 2. operating specifications -2 – Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 9

Advertising
background image

IM 12D08E02-01E

2-2 Specification

H. Display

: Custom liquid crystal display,

with a main display of 3

1

/

2

digits 12.5 mm high. Message

display of 6 alpha numeric

characters, 7 mm high. Warn-

ing flags and units (mS/cm,

kΩ·cm, µS/cm and MΩ·cm) as

appropriate.

I. Contact Outputs

- General : Four (4) SPDT relay contacts

with LED indicators. For S1,

S2, and S3, the LED is on

when relay power is removed.

NOTE: For S4 (FAIL) LED

lights when relay is

deenergised (Fail

safe).

Contact outputs configurable

for hysteresis and delay time.

- Switch capacity

: Maximum values 100 VA,

250 VAC, 5 Amps.

Maximum values 50 Watts,

250 VDC, 5 Amps.

- Status

: High/low process alarms,

selected from conductivity,

resistivity and temperature.

Contact output is also

available to signal “Hold

active”

- Control function

: On/Off

PI pulsed : Proportional duty cycle control

with integral term.

PI frequency : Proportional frequency control

with integral term.

(PI control on Conductivity/

Resistivity only) In addition

FAIL alarm for sys tem and

diagnostic errors on S4.

J. Power Supply

Supply voltage rating: 115, 230 VAC

Applicable range: 97.8 to 132.2, 195.5 to

264.5 VAC

Supply frequency rating: 50 / 60 Hz

Applicable range: 50 Hz ± 5% / 60 Hz ± 5%

Power consumption: Maximum 10 VA for

steady operation

K. Shipping Details

Package size

: W x H x D

290 x 300 x 290 mm.

11.5 x 11.8 x 11.5 in.

Packed weight

:approx. 2.5 kg (5lb).

2-2. Operating specifications

A. Performance : Conductivity

- Linearity : ± 0.5 % FS

- Repeatability : ± 0.5 % FS

- Accuracy : ± 0.5 FS

Performance : Resistivity

- Linearity : ± 0.5 FS

- Repeatability : ± 0.5 % FS

- Accuracy : ± 0.5 % FS

Performance : Temperature with Pt1000Ω,

Ni100Ω and PB36 NTC

- Linearity : ± 0.3 °C

- Repeatability : ± 0.3 °C

- Accuracy : ± 0.3 °C

Performance : Temperature with PT100Ω

and 8k55Ω

- Linearity : ± 0.4 °C

- Repeatability: ± 0.4 °C

- Accuracy : ± 0.4 °C

Note; The following tolerance are added to

above

performance.

mA output tolerance : ± 0.02 mA of

"0/4 - 20 mA"

Digital display tolerance: +1 digit

Performance : Temperature compensation

- NaCl table : ± 1 %

- Matrix

: ± 3 %

- Step response: 90 % (< 2 decades) in ≤ 6

seconds

B. Ambient operating temperature

: -10 to +55 °C (14 to 131 ºF)

C. Storage temperature

: -30 to +70 °C (-20 to 160 ºF)

D. Humidity

: 10 to 90% RH non-condensing

Advertising