Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 44

Advertising
background image

IM 12D08E02-01E

Parameter setting 5-19

Code Display

Function

Function detail

X

Y

Z

Default values

mA output functions
30

*mA

mA output range

mA1 = 0-20 mA

0

1.1

mA1 = 4-20 mA

1

4-20

mA2 = 0-20 mA

0

mA2 = 4-20 mA

1

4-20

31

*OUTP.F

mA output functions

Calculated value (Service Code 58)

0

0

1.1

Meas. conductivity linear

1

1

Cond.1/2

Meas. conductivity table

2

2

Meas. temperature

3

3

T1-T2

4 4

Direct/reverse action For rising signal, set for increasing mA out. 5

5

(direct)

(For PI control only)

For rising signal, set for decreasing mA out. 6

6

(reverse)

32

*BURN

Burn function

mA 1 No burnout

0

0.0 No Burn.

mA 1 Burnout downscale

1

mA 1 Burnout upscale

2

mA 2 No burnout

0

No Burn.

mA 2 Burnout downscale

1

mA 2 Burnout upscale

2

33

*RG.mA1(2) PI range

Proportional range for mA control signal

10 %

(use >, ^, ENT keys to adjust value)

34

*tI.mA1(2)

Integral time

100 sec.

(for PI control)

35

*TABL1

Output table for mA1

*0%

Linearization table for mA1 in 5% steps.

*5%

The measured value is set in the main

*10%

display using the >, ^, ENT keys, for

...

each of the 5% interval steps.

...

Where a value is not known, that value may

*90%

be skipped, and a linear interpolation will

*100%

take place.

36

*TABL2

Output table for mA2 Similar to code 35

37

*DAMP

Damping time

Damping on mA output 0-120 sec.

0. sec.

38-39

Not used

EXAMPLES:

A = bi-lineair

B = hyperbolic (2 decades)
C = logarithmic (2 decades)

D = linear

bi-lin = bi-lineair over 2 decades

log 2 = logarithmic over 2 decades

log 3 = logarithmic over 3 decades

hyp 2 = hyperbolic over 2 decades

hyp 3 = hyperbolic over 3 decades

NOTE: Multiply the values from the table with appropriate

factors to get the end-scale value you want.

% of output range

100

0 10 50 100 110

B

50

0

A

C

D

% of conductivity range

Fig. 5-2. Percentage of mA-output range versus

percentage of conductivity range

Table 5-4. Example of output tables

Output

bi-lin

log 2

log 3

hyp 2

hyp 3

0% 0.0 1.0 0.10 1.00 0.10

5% 1.0 1.3 0.14 1.20 0.27

10%

2.0

1.6

0.20

1.82

0.43

15%

3.0

2.0

0.28

1.90

0.61

20% 4.0 2.5 0.40 2.00 0.83

25%

5.0

3.2

0.56

3.75

1.10

30%

6.0

4.0

0.79

4.80

1.36

35%

7.0

5.0

1.12

5.92

1.68

40%

8.0

6.3

1.58

7.00

2.05

45% 9.0 7.9 2.24 8.31 2.49

50%

10.0

10.0

3.16

10.00

3.00

55%

20.0

12.6

4.47

11.85

3.66

60%

30.0

15.8

6.31

14.00

4.33

65%

40.0

20.0

8.91

16.65

5.22

70%

50.0

25.1

12.6

19.50

6.80

75%

60.0

31.6

17.8

23.80

8.25

80% 70.0 39.8 25.1 29.55 11.0

85%

80.0

50.1

35.5

36.70

14.8

90%

90.0

63.1

50.1

48.50

21.8

95%

100.0

79.4

70.8

68.60

36.5

100% 110.0 100.0 100.0 100.0 100.0