IM 12D08E02-01E

5-4 Parameter setting

5-2. Commissioning mode

5-2-1. Introduction

In order to obtain peak performance from the EXA DC402, you must set it up for each custom application.

Setpoints

Alarms are set by default

S1 - high process alarm

S2

-

low

process

alarm

S3

-

not

activated

S4 - Fail

The setpoints are at arbitrary default value. Therefore, you must set these to meaningful

values, or set them to off. (See service codes 40 to 49 and user interface codes 50 to 59.)

Output ranges mA output 1 is set as default to 0-100 µS/cm or 0-19.99 MΩ•cm.

For enhanced resolution in more stable measuring processes, it may be desirable to

select 5-10 µS/cm range, for example, and maybe 0-25 °C temperature range.

Service codes 30 to 39 can be used to choose other output parameters on mA output 2.

Choose from Table, temperature or PI control.

Hold

The EXA DC402G converter has the ability to “HOLD” the output during maintenance

periods. This parameter should be set up to hold the last measured value, or a fixed

value to suit the process.

Service

This selection provides access to the service menu.

What follows are pictorial descriptions of typical frontplate pushbutton sequences for each parameter set-

ting function. By following the simple YES/NO prompts and arrow keys, users can navigate through the

process of setting range, setpoints, hold and service functions.