3-4. ma output functions, 3-4. ma output functions -18 – Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 43

Advertising
background image

IM 12D08E02-01E

5-18 Parameter setting

0

1.000

800

600

400

200

0

100

80

60

40

20

0

2 4 6 8 10 12 14 16 18 20 22 24

5-3-4. mA output functions

Code 30

*mA

Select 4-20mA or 0-20mA according to associated equipment (recorders,

controllers etc.)

Code 31

*OUTP.F Note: For resistivity measurement, read resistivity in stead of conductivity.

Output mA1

Conductivity linear

(terminals 61&62) Conductivity with 21 point output table. (The table can be configured to

give an output linear to concentration, see example at the end of this

page).

Output mA2

Conductivity linear

(terminals 65&66) Conductivity with 21 point output table.

Temperature linear

PI control on conductivity (analog output control signal with proportional

and integral functions).

Direct or reverse action of the mA control output. Direct gives rising output with

rising measurement. Reverse gives falling output with rising measurement.

Code 32

*BURN

Diagnostic error messages can signal a problem by sending the output signals

upscale or downscale (22mA or 0/3.5mA). This is called upscale or downscale

burnout, from the analogy with thermocouple failure signalling of a burned-out

or open circuit sensor. In the case of the EXA the diagnostics are extensive

and cover the whole range of possible sensor faults.

Code 33

*RG.mA1(2) This function sets the proportional range for the mA output control signal. The

range setting is expressed in % of setpoint.

Code 34

*tI.mA1(2) This function sets the integral time for the mA output control signal

Code 35-36 *TABL1(2) The table function allows the configuration of an output curve by 21 steps

(intervals of 5%)

The following example shows how the table may be configured to linearize

the output with a W/W% curve. On the next page some other possibilities are

shown.

Code 37

*DAMP

Transmission signal damping (not mA control output).

Code mA

mA

Output

0-20

4-20

% H

2

SO

4

mS/cm

00

0 0

00.

4

00.0

0

00

0

00

5 1

0

4.8

0

1.25

0

60

0

10 2

0

5.6

00

2.5 113

0

15 3

0

6.4

0

3.75 180

0

20 4

0

7.2

00.0

5 218

0

25 5

00.

8

0

6.25

290

0

30

6

0

8.8

00

7.5 335

0

35 7

0

9.6

0

8.75 383

0

40 8 10.4

00.

10 424

0

45

9

11.2

11.25

466

0

50 10

0.

12

0

12.5 515

0

55 11 12.8 13.75

555

0

60

12

13.6

00.

15 590

0

65

13

14.4

16.25

625

0

70 14 15.2

0

17.5

655

0

75 15

0.

16

18.75

685

0

80

16

16.8

00.

20 718

0

85

17

17.6

21.25

735

0

90 18 18.4

0

22.5 755

0

95 19 19.2 23.75

775

100

20 20.0

00.

25 791

Fig. 5-1. Linearization of output

Example: 0-25% Sulfuric acid

Concentration (% by weight)

Output in %

Table 5-3.

Conductivity (mS/cm)

Conductivity

% Output