Dc402g specifications, 1. general specifications, Dc402g specifications -1 – Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 8: 1. general specifications -1

Advertising
background image

IM 12D08E02-01E

Specification 2-1

2-1. General specifications

A. Input specifications

: Two inputs , each 2-electrode

measurement with square wave

excitation, using cell constants(C)

from 0.008 to 50.0 cm

-1

, with up

to 60 metres (200ft) connection

cable.

B. Detection method

: Frequency, read-pulse position

and reference voltage are

dynamically optimized.

C. Input ranges

Minimum : 1µS x C at process tem perature

(underrange 0.000 µS/cm).

Maximum : 25 mS x C at process tem perature

(overrange 30 mS x C).

-Resistivity : 0.00 kΩ - 999 MΩ/C at 25 °C

(77 °F) reference temperature.

Minimum : 40 Ω/C at process temperature

(underrange 0.001 kΩ x cm).

Maximum : 1 MΩ/C at process tem pera ture

(overrange 999 MΩ x cm).

-Temperature

Pt1000

: -20 to +250 °C (0 to 500 °F)

Pt100 and Ni100

: -20 to +200 °C (0 to 400 °F)

8K55 NTC : -10 to +120 °C (10 to 250 °F)

PB36 NTC : -20 to +120 °C (0 to 250 °F)

D. Span

Conductivity/Resistivity

- Min. span : 0.010 µS/cm; 0.001 kΩ x cm

up to 90% maximum zero

suppression.

- Max. span : 1500 mS/cm; 999 MΩ x cm

Ratio (cell1/cell2)

- Min. span : 00.0

- Max. span : 19.99

Difference (cell1- cell2)

- Min. span : 0.010 µS/cm

- Max. span : 400 mS/cm

% Passage (100x[cell2/cell1] )

- Min. span : 00.0

- Max. span : 199.9

% Rejection (100x[( cell1-cell2)/cell1] )

- Min. span : 0.1

- Max. span : 400

% Deviation (100x[ (cell2-cell1)/cell1] )

- Min. span : 0.1

- Max. span: 400

VGB-directive 450 L

- Min. span : 1.0 pH

- Max. span : 14.0 pH

Temperature

- Min. span : 25 °C (50 °F)

- Max. span : 250°C (500 °F)

Difference Temperature

- Min. span : 25 °C (50 °F)

- Max. span : 250 °C (500 °F)

E. Transmission Signals

: Two isolated outputs of 0/4-20

mA DC with common negative.

Max. load : 600 Ω.

Auxiliary output can be chosen

from conductivity, linearized

conductivity, resistivity,

temperature, differential

temperature calculated value

or PI control of conductivity/

resistivity.

Burn up (22 mA) or Burn down

(0/3.5 mA) to signal failure.

F. Temperature compensation

: Automatic, for temperature

ranges mentioned under C

(input ranges).

- Reference temperature

: programmable from 0 to 100 °C

or 30 to 210 °F (default 25 °C).

G. Compensation algorithm

: According IEC 60746-3

NaCl tables (default). Two

independent user programmable

temperature coefficients, from

0% to 3.5% per °C (°F) by

adjustment or calibration.

- Matrix compensation

: with conductivity function

of concen-tration and

temperature. Choice of 5

preprogrammed matrixes

and a 25-points user-

programmable matrix.

2. DC402G SPECIFICATIONS

Advertising