Parameter setting, 1. maintenance mode, 1-1. introduction – Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 26: Parameter setting -1, 1. maintenance mode -1, 1-1. introduction -1

Advertising
background image

IM 12D08E02-01E

Parameter setting 5-1

5. PARAMETER SETTING

5-1. Maintenance mode

5-1-1. Introduction

Standard operation of the EXA instrument involves use of the Maintenance (or operating) mode to set up

some of the parameters.

Access to the Maintenance mode is available via the six keys that can be pressed through the flexible

window in the instrument front cover. Press the “MODE” key once to enter this dialog mode.

(Note that at this stage the user will be prompted for a pass code where this has been previously set up

in service code 52, section 5.)

Setpoint

Select and adjust setpoint (when enabled in service menu section 5-3, service code 51).

See adjustment procedure 5-2-2.

Calibrate

See “calibration” section 6.

Display setting See “operation” section 4.

Hold

Manually switch on/off “hold” (when enabled in commissioning menu). See adjustment

procedure 5-2-4.