Function, Command/response (for normal operation), Example – Yokogawa JUXTA M Series Digital Limit Alarms MVTK User Manual

Page 25

Advertising
background image

<Toc> <Ind>

<3. PC Link Communication >

3-11

IM 77J04J11-01E 1st Edition : 2006.08.25-00

BRS Specifies I relays to be monitored on a bit-by-bit basis

Function

Specifies the I-relay numbers to be monitored on a bit-by-bit basis. Note that this com-
mand simply specifies I relays. Actual monitoring is performed by the BRM command after
the I relay numbers are specified by this command.

When the volume of data is large and you wish to increase the baud rate, it is effective to
use a combination of the BRS and BRM commands rather than just the BRR command.

If the power supply is turned off, the specified I-relay numbers will be erased.

The number of I relays to be specified at a time is 1 to 32.

For the format of response in the event of failure, see subsection 3.1.2.

The command shown below includes the checksum function.
When performing communication without checksum, do not include the 2-byte
checksum element in the command.

Command/Response (for normal operation)

Number of Bytes

1

2

2

1

3

2

5

1

5

1

Command

element

STX

Address

number

(ADR)

CPU

number

01

0

Number

of bits

(n)

I relay

number 1

BRS

Comma

or space

I relay

number 2

Comma

or space

Command (continued)

• • •

5

2

1

1

• • •

I relay

number n

Checksum

ETX

CR

1

2

2

2

2

1

1

Number of Bytes

Response

element

STX

Address

number

(ADR)

CPU

number

01

OK

Checksum

ETX

CR

Example:

Specifying that the burnout, alarm 1 and alarm 2 of the M Series with address number 01
are to be monitored.

(This command simply specifies the registers to be monitored.)

[Command]

[STX]01010BRS03I0007,I0001,I0002B9[ETX][CR]

OK is returned in response to the above command.

[Response]

[STX]0101OK5C[ETX][CR]