Ocp setting (:curr:prot) – GW Instek PSB-2000 Series User Manual User Manual
Page 114

PSB-2000 Series User Manual
114
The <value> range for single-unit operation is as shown
below:
PSB-2400L & PSB-2400L2: 0.00 to 41.00
PSB-2800L: 0.00 to 82.00
PSB-2400H: 0.00 to 3.07
PSB-2800H: 0.00 to 6.15
Two decimal places are valid.
The setting range changes when the power supplies are
used in parallel.
Resolution in parallel operation: 100mA
Setting range: 0.0 to (total current x 1.025)
Application
example
:CURR 10.10
Sets the output current of a single channel model to
10.10A.
Query
:CURR?
Inquires about the set output current of a single channel
model.
:CURR:A?
Inquires about the set output current of CH1.
:CURR:B?
Inquires about the set output current of CH2.
Response
example
10.10
Indicates that the set output current is 10.10A.
Remark
When the tracking mode is on, setting CH1 will affect
CH2. Setting CH2 in the tracking mode will cause an
execution error.
OCP setting (:CURR:PROT)
This command sets or inquires about the OCP (over-current
protection) value.
Setting
:CURR:PROT <value>
Sets the OCP value of a single channel model.
:CURR:PROT:A <value>
Sets the OCP value of CH1.
:CURR:PROT:B <value>
Sets the OCP value of CH2.