EXTERNAL CONTROL

121

External Control of the Current Output ON/OFF
(:EXT:CURR)

This command chooses whether the external control is used to
control the output current or if the front panel controls are used to
control the output current.

Setting

:EXT:CURR <value>
Activates or deactivates external control of the output
current of a single channel model.
:EXT:CURR:A <value>
Activates or deactivates external control of the output
current of CH1.
:EXT:CURR:B <value>
Activates or deactivates external control of the output
current of CH2.
<value> is as shown below:
0: Deactivates external control of the output current
(Allows the current output to be controlled with the front
panel controls or via remote control).
1: Activates external control of the current output.

Application
example

:EXT:CURR 1
Selects the external control of the current output.

Query

:EXT:CURR?
Inquires about the setting.

Response
example

1
Indicates that external control of the current output is
enabled. The query return value has the same meaning as
the setting <value>.

Remark

This command causes an error during sequence
operation (RUN).

Output switching (:EXT:OUTP)

This command selects output On/Off with external contacts.

Setting

:EXT:OUTP <value>
<value> is as shown below:
0: Control from the panel or through communication
1: Output On/Off with external contacts

Application
example

:EXT:OUTP 1
Selects output On/Off with external contacts.

Query

:EXT:OUTP?
Inquires about the setting.