400 series, En - 20, The previous anomalous results cannot be saved – HT instruments SPEED418 User Manual

Page 21

Advertising
background image

400

Series

EN - 20

5. If the instrument detects a

voltage value higher than
10V at the input leads, the
screen reported here to the
side is displayed

L O W

- - - -

R +

R -

- - - -

 - - - - 

- - - m A

- - - m A

Vin > Vlim

AUTO

4.00

0.21

Func

Lim CAL

6. In case it was detected that

the calibrated resistance is
higher than the measured
resistance increased by
0.05

 (R

CAL

>R

MEAS

+0.05

),

the instrument gives a long
acoustic signal and displays
a screen similar to the one
reported here to the side

L O W

0 . 0 0

R +

R -

0 . 0 0

 0 . 0 0 

2 1 4 m A

2 1 4 m A

CAL > RES

AUTO

4.00

0.21

Func

Lim CAL

7. By using the CAL mode, if

the instrument detects a
resistance higher than 5

 at

input leads, a screen similar
to the one reported here to
the side is displayed, and the
instrument remains in the
condition of no resistance
calibrated

L O W

- - - -

R +

R -

- - - -

 - - - - 

- - - m A

- - - m A

Reset value

CAL

4.00

0.21

Func

Lim CAL

8. By using the CAL mode,

verifying the calibrated value
at the end of the CAL
procedure, if the condition:
R

CAL

≤ R

MEAS

≤ R

CAL

+ 0.05Ω

is not met, a screen similar
to the one reported here to
the side is displayed, and the
instrument remains in the
condition of no resistance
calibrated

L O W

1 . 9 8

R +

R -

1 . 9 8

 1 . 9 8 

2 1 0 m A

2 1 0 m A

Not correct

CAL

4.00

----

Func

Lim CAL

9.

The previous anomalous results cannot be saved