Fig. 18: Instrument connection for 400V + N + PE three-phase P-P line impedance

measurement by means of single cables and remote probe

Fig. 19: Instrument connection for 400V + PE (no N) three-phase P-PE fault loop

impedance measurement through single cables and remote probe

1.

Press the MENU key, move the cursor to
LOOP

in the main menu by means of the

arrow keys (,) and confirm with ENTER.
Subsequently the instrument displays a screen
similar to the one reported here to the side.

L O O P

----

- - - - A

FRQ=50.0Hz
VP-N=228V

VP-Pe=228V

P-PE STD

Func

Mod.

2.

Use the ,  keys to select the parameter to be modified, and the ,  keys
to modify the parameter value.
It is not necessary to confirm the selection with ENTER

.

Func

The virtual Func key allows setting the measuring mode of the

instrument, which may be: P-N, P-P, P-PE

UL

The virtual UL key, active only when an IT electrical system and the
P-PE mode have been set (§ 5.2.5), allows setting the limit value of
contact voltage for the system being tested, which may be: 50V,
25V

Mod.

The virtual UL key, not active when an IT electrical system and the
P-PE mode have been set (§ 5.2.5), allows setting the instrument’s
operating mode, which can be: STD, Z2

ICAL

The virtual ICAL key, active only when the Z2Ω mode has been
selected with the MOD key, allows selecting the assumed short-
circuit or fault current displayed. Following values are available:
IkMax3Ph, IkMin3Ph, IkMax2Ph, IkMin2Ph, IkMaxP-N, IkMinP-
N, IkMaxP-PE, IkMinP-PE, IkSTD