Auxiliary measurements – HT instruments SPEED418 User Manual

Page 53

Advertising
background image

400

Series

EN - 52

7. AUXILIARY MEASUREMENTS

7.1.

AUX:

REAL TIME MEASUREMENT OF THE ENVIRONMENTAL
PARAMETERS

By means of external transducers, this function allows measuring the following
environmental parameters:

AIR

air speed by means of air-speed transducer

RH

air humidity by means of humidity transducer

TMP °F air temperature in °F by means of thermometric transducer
TMP °C air temperature in °C by means of thermometric transducer
dB

acoustic pressure by means of sound level transducer

Lux

lighting by means of luxmetric transducer

VOLT

input voltage (without applying any transduction constant)

Fig. 28: Connection between an

environmental probe and the instrument

Fig. 29: Connection between the sound level

probe and the instrument

1.

Press the MENU key, move the cursor to AUX
in the main menu by means of the arrow keys
(,) and confirm with ENTER. Subsequently
the instrument displays a screen similar to the
one reported here to the side. Except in dB
mode, the instrument measures and shows on
the display the instant value of the input
parameter in real time, and RUN appears in
the right bottom part of the display

A U X

I n 1 =

7 . 0 8 L u x


Lux LX

20 RUN

Func

FS

2.

Use the ,  keys to select the parameter to be modified, and the ,  keys
to modify the parameter value.
It is not necessary to confirm the selection with ENTER

.

Func

The virtual Func key allows setting the the type of environmental

measurement to perform, which may be: AIR, RH, TMP °F, TMP
°C, dB, Lux, VOLT

FS

The virtual FS key, active only when the Lux mode has been
selected with the Func key, allows selecting the clamp full scale
among the values: 20, 2k, 20k