Badger Meter Vortex Meters User Manual
Page 11
Page 9
Exiting Edit Mode (Or putting the meter in-service) - To put the meter in-service, click the
“Edit/In-Service” pull-down menu on the Main screen and select “Put Meter In-Service”. The
meter must be put back in-service after making changes to the meter. Otherwise, the analog
output of the meter will not be updated.
Racine Federated HART Interface Utility
FL_METER
0
4
0
72
0
RWG15
Tag:
PV (Volume)
Model Number:
Analog Output Current:
Percent of Range:
Process Temperature:
Process Pressure:
ft3/min
mA
%
degF
psi
Connect
Connect
Update
Begin Continuous Display Update
Tools
About
Com Port
Save/Restore Configuration
Configure Meter
Continuous Screen Update:
Fixed Current Value:
Password Protection:
Configuration Changed, Not Saved:
Edit/In-Service
Edit/In-Service
Put Meter In Service
Enter Edit Mode
Enter Edit Mode
Meter In-Service
Indicator
Edit Mode
Indicator
Meter In-Service:
Edit Mode:
power up state - Meters containing embedded software version 11 or higher will always
power up “In-Service”. Meters containing earlier embedded software versions will power up in
the same state that they were powered off in.