The active alarm current setting is read from the meter and displayed when this dialog box is

opened. To change the setting, click the desired selection and then click the Write button.

Alarm Selection

High (21.75 mA)

Low (3.75 mA)

Alarm Current Level:

Write

Read

Close

NOTE: Save the changes to EEPROM and put the meter in-service after making changes.

pipe iD

The internal diameter of the pipe or the pipe ID can be read from the “Pipe ID” dialog box. To

bring up this box, click the “Configure Meter” pull-down menu on the Main screen and then

select “Pipe ID”.

For insertion style meters, this box allows the user to set the inside pipe diameter to the appro-

priate size. The pipe ID is used to calculate volumetric, mass, and standard flow.

Pipe ID

Pipe ID:

in.

1.375

Write

Read

Close

For wafer style meters, the pipe ID cannot be modified and the Write button is disabled.

NOTE: Save the changes to EEPROM and put the meter in-service after making changes.

low flow Cutoff

The low flow cutoff can be read and configured from the “Low Flow Cutoff” dialog box. To bring

up this box, click the “Configure Meter” pull-down menu on the Main screen and then select

“Low Flow Cutoff”.

Low Flow Cutoff

Low Flow Cutoff

Hz

25

Write

Read

Close

Page 29