Output Current and Percent of Range - This selection causes the analog output current and

the percent of range values to be output during burst mode. Command number 2 of the Racine

HART Message Specification defines the format of this message.

output Current and four Dynamic variables - This selection causes the analog output

current, the primary variable (PV), the vortex shedding frequency, the process temperature,

and the process pressure to be output during burst mode. Command number 3 of the Racine

HART Message Specification defines the format of this message.

NOTE: Burst mode slows HART communication and will cause the Racine HART Interface to take longer to

read and write to the meter. Turn burst mode off when it is not being used.

NOTE: Save the changes to EEPROM and put the meter in-service after making changes

standard flow

The base process conditions for the standard flow calculations can be configured from the

“Standard Flow” dialog box. To bring up this box, click the “Configure Meter” pull-down menu

on the Main screen and then select “Standard Flow”.

Standard Flow

Base Pressure:

Base Temperature:

psi

°F

14.696

60

Write

Read

Close

The data displayed by this window is read from the meter when the dialog box is opened and

when the user clicks the Read or Write buttons

The “Base Temperature” and “Base Pressure” values used to calculate the standard (or nor-

malized) flow, can be read and modified in this dialog box.

NOTE: Save the changes to EEPROM and put the meter in-service after making changes.

password protection

The password protection feature prevents unauthorized users from editing the meter’s settings.

Users are prohibited from putting the meter in “Edit Mode” without entering a correct password

when “Password Protection” is enabled.

Password protection is enabled and disabled using the “Password Protection” dialog box. To

bring up this box, click the “Configure Meter” pull-down menu on the Main screen and then se-

lect “Password Protection”. The password protection feature can only be enabled and disabled

when the meter is on-line.