Badger Meter Vortex Meters User Manual

Page 30

Advertising
background image

Password Protection

Password Verification:

Password:

Write

Read

Close

Password Protection Enabled

To password protect the meter, click the “Password Protection Enabled” box and enter a pass-

word into the “Password” box. The password must be repeated in the “Password Verification”

box before the “Write” button is clicked. There is a “Password Protection” status indicator on

the main screen. The indicator is green when password protection is enabled and black when

password protection is disabled.

In order to configure the meter, password protection must be disabled. There are two ways to

disable password protection. One way is to use the “Password Protection” dialog box to un-

check the “Password Protection Enabled” box and type in the correct password in the “Pass-

word” field and then click “Write”. The other way to disable password protection is to simply

attempt to edit the meters settings and the user will be prompted for the password.

Forgot password? - If the user forgets the password, the

password protection can be disabled by removing the se-

curity jumper from the meter’s circuit board. To remove the

jumper, first remove the end-cap that covers the electronic

circuit boards. If the model has a display, remove the dis-

play-surround in order to remove the security jumper. The

security jumper is the only jumper on the circuit board so

it is easy to find. With the security jumper removed, the password protection can be disabled

using one of the two methods mentioned above. Replace the security jumper to once again

enable password protection.

alarm selection

If a failure is detected in the meter, the loop current will go to an alarm value. The alarm current

may be configured to be either “Low” (3.75 mA) or “High” (21.75 mA).

Do not confuse this feature with process alarms that indicate that a particular level of flow was

exceeded or not met. This feature only reflects the operating condition of the electronics and

sensory apparatus in the meter.

The alarm current can be configured from the “Alarm Selection” dialog box. To bring up this

box, click the “Configure Meter” pull-down menu on the Main screen and then select “Alarm

Selection”.

Page 28

Security

Jumper