Capacitance Compensation

Capacitance Compensation

When the capacitance/conductance measurement is performed through the Agilent
B2200, LCR meter measures the capacitance/conductance of the path including a
device under test (DUT), matrix switches, extension cables and so on. So, the data
measured by the LCR meter is far from the DUT’s capacitance/conductance.

The Agilent B2200 VXIplug&play driver provides the functions used to
compensate the capacitance/conductance measured by the Agilent 4284A LCR
meter in the measurement environments described in “Required Conditions” on
page 4-21.

This section explains how to use the capacitance compensation function.

•

“Capacitance Compensation Function”

•

“Required Conditions”

•

“To Create Compensation Data File”

•

“To Perform Measurement and Compensation”

Capacitance Compensation Function

Driver functions used for the capacitance compensation are listed below.

•

agb220xa_selectCompenFile function

•

agb220xa_compenC function

NOTE

Corrected data by the function is not guaranteed. But typical data (supplemental
data) is as follows.

Capacitance measurement accuracy (typical): ±1 % ±0.5 pF

This typical data is for the following measurement conditions:

Measurement frequency: 1 kHz to 1 MHz

Measurement range:

Maximum 1000 pF

Measurement terminal:

At the end of the Agilent 16494A/B/C cable connected
to the switch module output terminals.

The typical data does not apply to anything extended from the 16494A/B/C cable.
The conditions described in “Required Conditions” on page 4-21 must be satisfied.