Agilent Technologies Agilent B2201A User Manual

Page 53

Advertising
background image

Agilent B2200 User’s Guide, Edition 4

2-17

Installation

Output Connections

The 4155/4156 semiconductor parameter analyzer's Intlk connector provides the
interlock signal and a LED drive signal. If a LED is connected between pin 4 and
pin 5 (or 6) of the interlock connector, the LED lights to indicate high voltage output
when more than ±40 V is forced from an SMU in the 4155/4156.

Figure 2-6

Dimensions of LED (Agilent part number 1450-0641)

To Connect Interlock Circuit to Instrument

Before beginning the measurement, connect the interlock circuit to the instrument’s
interlock connector as follows.

•

For the instruments which has a BNC-Type interlock connector:

1. Get the following parts.

•

Agilent 16493J Interlock cable, 1 ea.

•

Agilent 16435A Interlock cable adapter, 1 ea.

2. Connect the 16493J interlock cable between the interlock circuit and the

16435A adapter.

3. Connect the BNC cable (furnished with the adapter) between the adapter and

the instrument’s interlock connector.

•

For Agilent 4155/4156/E5260/E5270:

Connect the 16493J interlock cable between the interlock circuit and the
instrument’s interlock connector. You can connect it directly without using any
adapter.

6

10

11

5

5.6

5

C

a

th

od

e(

-)

A

n

od

e(

+

)

Units: mm