Switch modules – Agilent Technologies Agilent B2201A User Manual
Page 30

1-12
Agilent B2200 User’s Guide, Edition 4
Introduction
Specifications
Switch Modules
Table 1-1 lists the specifications when the switch module is installed in the
mainframe. In the table, the conditions are as follows:
Conditions: 23
°C ± 5 °C, 5 % to 60 % R.H.
Table 1-1
Agilent B2210A/B2211A Switch Module Specifications
Agilent B2200A mainframe supports Agilent B2210A module.
Agilent B2201A mainframe supports Agilent B2211A module.
Mixed configuration of the switch modules is not supported.
B2210A B2211A
Maximum current rating (A)
I-V port
1.0
1.0
AUX port
0.5
0.5
Maximum voltage rating (V)
I-V port, channel to common
200
200
I-V port, channel to other channel
300
300
AUX port, channel to common
100
100
AUX port, channel to other channel
100
100
Closed channel residual resistance (
Ω)
I-V port
0.6
0.6
AUX port
1.5
1.5
Channel isolation resistance (
Ω)
I-V port
10
14
5
× 10
13
AUX port
10
9
10
9