Agilent Technologies Agilent B2201A User Manual
Page 124

4-24
Agilent B2200 User’s Guide, Edition 4
Programming
Capacitance Compensation
Table 4-9
Template Compensation Data Files
File name
a
a. <path>: driver_install_folder\AGB220XA\ccdata (e.g. C:\temp\AGB220XA\ccdata)
Measurement environment that template targets
Switch
module
Cable
b
b. Model number of the cable connected between the switch module and the DUT interface.
DUT
interface
c
c. Agilent B2220A probe card interface or Agilent 16495F/G connector plate.
Coefficients to
be modified
<path>\B2210A\pcif\triax\3m.data
B2210A
16494A-002
B2220A
C3H and C3L
<path>\B2210A\pcif\triax\4m.data
16494A-005
<path>\B2210A\pcif\kelvin\3m.data
16494C-002
<path>\B2210A\pcif\kelvin\4m.data
16494C-005
<path>\B2210A\cable\triax\1_5m.data
16494A-001
16495F/G
C2H, C2L,
C3H, and C3L
<path>\B2210A\cable\triax\3m.data
16494A-002
<path>\B2210A\cable\triax\4m.data
16494A-005
<path>\B2210A\cable\kelvin\1_5m.data
16494B-001
<path>\B2210A\cable\kelvin\3m.data
16494B-002
<path>\B2210A\cable\kelvin\4m.data
16494C-005
<path>\B2211A\pcif\triax\3m.data
B2211A
16494A-002
B2220A
C3H and C3L
<path>\B2211A\pcif\triax\4m.data
16494A-005
<path>\B2211A\pcif\kelvin\3m.data
16494C-002
<path>\B2211A\pcif\kelvin\4m.data
16494C-005
<path>\B2211A\cable\triax\1_5m.data
16494A-001
16495F/G
C2H, C2L,
C3H, and C3L
<path>\B2211A\cable\triax\3m.data
16494A-002
<path>\B2211A\cable\triax\4m.data
16494A-005
<path>\B2211A\cable\kelvin\1_5m.data
16494B-001
<path>\B2211A\cable\kelvin\3m.data
16494B-002
<path>\B2211A\cable\kelvin\4m.data
16494C-005