Front Panel
Front Panel
The Agilent B2200 series provides the front panel keys, the LCD, and the LED
matrix display for the status monitor and connection setup.
•
Line switch
Used to turn the Agilent B2200 on or off.
•
LCD
Used to monitor the status and set the connection. See Chapter 3 for the details.
•
Front panel keys
Used to set the Agilent B2200. See Chapter 3 for the details.
•
LED matrix display
Displays the status of the matrix switches. Also used to set the matrix connection
with the light pen.
•
SMU Input
Inputs for the source monitor unit (SMU). Eight input ports. Up to four kelvin
inputs are available. Triaxial BNC connector.
•
AUX Input
Multipurpose inputs. Six input ports. Coaxial BNC connector. CMH and CML
terminals are the input ports for the capacitance measurement.
•
Light Pen
Connector for the Agilent 16443A Light pen. Used to set the matrix connection
with the LED matrix display.
CAUTION
For the SMU Input terminals, the maximum measurement voltage/current/voltage
between terminals are ±200 V/1 A/300 V, respectively. And for the AUX Input
terminals, they are ±100 V/0.5 A/100 V.
Do not apply an input signal over these limits to the input terminals. If you do, the
Agilent B2200 will be damaged. If you use a bias source that has current limit
capability, set the bias source current limit.