Route]:couple:port:detect – Agilent Technologies Agilent B2201A User Manual

Page 164

Advertising
background image

5-36

Agilent B2200 User’s Guide, Edition 4

SCPI Command Reference
[:ROUTe]:COUPle:PORT:DETect

At *RST, no input ports are coupled.

Syntax

[:ROUTe]:COUPle:PORT card_number,'couple_port'

[:ROUTe]:COUPle:PORT? card_number

Query response

port_number{, port_number} <newline><^END>

port_number = 1 or 3 or 5 or 7 or 9 or 11 or 13

Example

OUTPUT @Agb2200;":ROUT:COUP:PORT ALL,'1,3'"

OUTPUT @Agb2200;":ROUT:COUP:PORT? 1"

ENTER @Agb2200;A$
In this example, A$ will be 1,3.

[:ROUTe]:COUPle:PORT:DETect

This command detects the input ports connected to the Kelvin cable, and assigns
them as the input couple ports that will be used for the Kelvin connection. The input
couple port setting is effective for the all cards. This command overwrites the
previous couple port setting.

The couple mode is set by “[:ROUTe]:COUPle[:STATe]”. The couple port setting
will be cleared and updated by “[:ROUTe]:COUPle:PORT”.

The input port numbers of the couple ports can be got by
[:ROUTe]:COUPle:PORT?.

After this command is executed, the all relay settings will be the same as after the
:ROUT:OPEN:CARD ALL command execution.

Syntax

[:ROUTe]:COUPle:PORT:DETect

Example

OUTPUT @Agb2200;":ROUT:COUP:PORT:DET"

Parameter

Explanation

card_number

For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL
ALL is not available for query.

couple_port

One or more input port numbers: 1, 3, 5, 7, 9, 11 or 13.

Enclose by single quotation marks. Separate multiple input
port numbers by comma. For example: '1,5'

Advertising