Diagnostic:test:frame:clear, Diagnostic:test:frame[:execute – Agilent Technologies Agilent B2201A User Manual

Page 150

Advertising
background image

5-22

Agilent B2200 User’s Guide, Edition 4

SCPI Command Reference
:DIAGnostic:TEST:FRAMe:CLEar

Syntax

:DIAGnostic:TEST:CARD:STATe? card_number

Query response

test_result <newline><^END>

1: fail
0: pass
-1: not tested

Example

OUTPUT @Agb2200;":DIAG:TEST:CARD:STAT? 1"

ENTER @Agb2200;A

:DIAGnostic:TEST:FRAMe:CLEar

This command clears test result of the specified B2200 test.

Syntax

:DIAGnostic:TEST:FRAMe:CLEar CONTroller | FPANel | LED |
PEN | BEEPer

Example

OUTPUT @Agb2200;":DIAG:TEST:FRAM:CLE CONT"

:DIAGnostic:TEST:FRAMe[:EXECute]?

This command executes the specified B2200 test, then returns the test result. See
“Selftest Menu” on page 3-34 to perform the test.

After the controller test, the B2200 status becomes same as after *RST command
execution.

Parameter Explanation

card_number

card number: 1, 2, 3, 4

Parameter Explanation

CONTroller

Controller test

FPANel

Front panel interface test

LED

LED matrix test

PEN

Light pen test

BEEPer

Beeper test

Advertising