Safety summary – Agilent Technologies Agilent B2201A User Manual

Page 6

Advertising
background image

Safety Summary

The following general safety precautions must be observed during all phases of

operation, service, and repair of this instrument. Failure to comply with these

precautions or with specific warnings elsewhere in this manual may impair the

protections provided by the equipment. In addition, it violates safety standards of

design, manufacture, and intended use of the instrument. Agilent Technologies, Inc.

assumes no liability for customer’s failure to comply with these requirements.

NOTE

Agilent B2200 complies with INSTALLATION CATEGORY II for mains input and

INSTALLATION CATEGORY I for measurement input terminals, and POLLUTION

DEGREE 2 defined in IEC 61010-1.
Agilent B2200 is INDOOR USE products.

GROUND THE INSTRUMENT

This is Safety Class I instrument. To minimize shock hazard, the instrument

chassis and cabinet must be connected to an electrical ground. The power

terminal and the power cable must meet International Electrotechnical

Commission (IEC) safety standards.

DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE

Do not operate the instrument in the presence of flammable gases or fumes.

Operation of any electrical instrument in such an environment constitutes a

definite safety hazard.

KEEP AWAY FROM LIVE CIRCUITS

Operation personnel must not remove instrument covers. Component replacement

and internal adjustments must be made by qualified maintenance personnel. Do

not replace components with power cable connected. Under certain conditions,

dangerous voltages may exist even with the power cable removed. To avoid

injuries, always disconnect power and discharge circuits before touching them.

DO NOT SERVICE OR ADJUST ALONE

Do not attempt internal service or adjustment unless another person, capable of

rendering first aid and resuscitation, is present.

DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT

Advertising