Output queue – Agilent Technologies Agilent B2201A User Manual

Page 185

Advertising
background image

Agilent B2200 User’s Guide, Edition 4

5-57

SCPI Command Reference

Output Queue

Output Queue

The Output Queue stores response messages until they are read. If an unread
message exists, Bit4 (Message Available

− MAV) of the Status Byte Register is set

to 1. So, Bit4 is used to synchronize information exchange with the controller. See
Figure 5-5.

Figure 5-5

Output Queue of B2200

The Response Formatter places Data Byte Messages and END messages into the
Output Queue in response to query commands. These messages are removed from
the Output Queue as they are read by the controller. As long as the Output Queue
contains an unread message, MAV is 1.

Advertising