Manual calibration – Vaisala Hydromet Data Collection Platform User Manual

Page 76

Advertising
background image

User's Guide ______________________________________________________________________

74 ___________________________________________________________________M210784EN-E

entering 24 in the Value field will change the offset of the
measurement to be 1 °C.

Manual Calibration

1.

On the Settings menu, select Calibration.

2.

The list of measurements available for calibration is displayed.
Select the measurement you want to calibrate from the list.

3.

Select Manual Calibration. The Manual Calibration window
shown in

Figure 38 on page 75

is displayed with information on the

measurement and its calibration. The Offset field shows the offset
for the measurement. The Factor field shows the gain of the
measurement; that is, the slope of the conversion curve.

Figure 38 on page 75

illustrates radiation calibration. Radiation

sensors tend to have individual characteristics; therefore, they
always need to be calibrated prior to operation.

NOTE

The measurement values obtained from the sensors in the Offset
Calibration
view are not updated automatically. To get the latest
measurement values, select the Refresh button.

The values you have changed in the Offset Calibration view but have
not yet sent to the QML logger by selecting Calibrate are displayed in
bold.