Format measurement output – Vaisala MMT162 User Manual

Page 32

Advertising
background image

User's Guide ______________________________________________________________________

30 __________________________________________________________________ M210934EN-C

Format Measurement Output

Use the serial line command FORM to change the format or select

certain parameters for the output commands SEND and R.

FORM [x]<cr>

where
x = Formatter string

Formatter string consists of parameters and modifiers. The MMT162

measures the following parameters:

- water activity (aw)
- temperature (T) (metric unit: ºC, non metric unit:ºF)
- ppm for transformer oil only (H

2

O)

When selecting the parameter, use its abbreviation (for example, aw).

The modifiers are presented in Table 6 on page 31.

Example (configure transmitter to output water activity and temperature):

form "aw=" 6.4 aw #t "t=" 6.2 t #r#n

OK

send

aw= 0.2644 t= 25.50

Example (configure transmitter to output ppm only):

form "Oil ppm= " h2o " " u3 #r#n

OK

send

Oil ppm= 16.6 ppm

Example (configure transmitter to default output format):

form /

OK

send

aw= 0.087 T= 24.0 'C

Advertising