Using mi70 – Vaisala MMT162 User Manual

Page 59

Advertising
background image

Chapter 6 ___________________________________________________ Calibration and Adjustment

VAISALA _______________________________________________________________________ 57

Using MI70

To adjust temperature:

1.

Connect MI70 to MMT162 via connector II.

2.

Turn on MI70.

3.

On the MI70 display, go to MenuFunctionsAdjustments.

4.

Follow the instructions given on the MI70 display.

5.

Select T from the list and press Select.