Chapter 5 ______________________________________________________________ Maintenance

VAISALA _______________________________________________________________________ 49

Table 9

Error Messages

Error Message

Action

Voltage error

Internal error or invalid supply voltage.

T meas error

Internal error, can be caused by

damage to the temperature sensor.

F meas error

Internal error, can be caused by

damage to the HUMICAP® sensor.

Voltage too low error

Supplied operating voltage is too low

for reliable operation.

Voltage too low for mA output

Supplied operating voltage is too low

for reliable operation of the analog

current output.

Voltage is too low for V output

Supplied operating voltage is too low

for reliable operation of the analog

voltage output.

Program flash check sum error

Internal error

Parameter check sum error

Internal error

INFOA check sum error

Internal error

SCOEFS check sum error

Internal error

CURRENT check sum error

Internal error

Unknown Serial Settings

If you are not sure what the current serial settings of your transmitter are,

follow the steps below to connect to the transmitter. This procedure

assumes you have the USB service cable available.

For more information on the USB service cable and the PuTTY terminal

application, see section Serial Communication on page 22.

1.

Connect the USB service cable to your PC, but do not connect it to

the transmitter yet.

2.

Start a terminal application and open a terminal session using the

default serial settings of 19200 8 N 1.

3.

Plug the USB service cable into the transmitter, and quickly send at

least one hash character # to the serial line:

#

4.

The normal startup of the transmitter is now interrupted, and the

transmitter is available for commands with the default serial

settings. You can now check the transmitter’s information by

issuing the ? command:

?

The transmitter’s currently stored serial settings are visible on the

line marked Baud P D S.